FDA Medical Device Recalls (openFDA)
Tosoh Bioscience Inc: ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocysteine in human serum, heparinized plasma or EDT A plasma on TOSOH AIA System Analyzers. Homocysteine measurements are used in the diagnosis and treatment of hyperhomocysteinemia or homocysteinuria.
- Recall number
- Z-0675-2019
- Recalling firm
- Tosoh Bioscience Inc
- Product
- ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocysteine in human serum, heparinized plasma or EDT A plasma on TOSOH AIA System Analyzers. Homocysteine measurements are used in the diagnosis and treatment of hyperhomocysteinemia or homocysteinuria.
- Status
- Terminated
- Initiated
- 2018-11-30
- Posted
- not on file
- Terminated
- 2021-01-14
- Reason
- not on file
- Root cause
- Device Design
- Action
- On November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiography. " Continue to use the Tosoh AIA assays listed in notice only for patients who have not undergone a fluorescein angiography procedure exam, or on patient samples which have been taken prior to receiving the fluorescein angiography procedure. " Complete & return the Urgent Medical Device Recall Acknowledgement Form within 15-days of receiving the notification. " Maintain a copy of the notification with your laboratory records and forward this information to others who may have received this product. Should you have any questions, please contact Bernadette OConnell at 1(800) 248-6764 or by email at Bernadette.oconnell@tosoh.com. She will be able to answer any questions Monday-Friday from 9 AM to 5 PM (PST).
- Quantity
- 992 units
- Distribution
- US Nationwide
- Lot, serial or model codes
- all lots
- 510(k) numbers
- ["K121533"]
- PMA numbers
- not on file
- Product code
- LPS
- Firm FEI
- 3005529799
- Firm city
- Grove City
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as