Skip to the content

FDA Medical Device Recalls (openFDA)

Tosoh Bioscience Inc: ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocysteine in human serum, heparinized plasma or EDT A plasma on TOSOH AIA System Analyzers. Homocysteine measurements are used in the diagnosis and treatment of hyperhomocysteinemia or homocysteinuria.

Recall number
Z-0675-2019
Recalling firm
Tosoh Bioscience Inc
Product
ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocysteine in human serum, heparinized plasma or EDT A plasma on TOSOH AIA System Analyzers. Homocysteine measurements are used in the diagnosis and treatment of hyperhomocysteinemia or homocysteinuria.
Status
Terminated
Initiated
2018-11-30
Posted
not on file
Terminated
2021-01-14
Reason
not on file
Root cause
Device Design
Action
On November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiography. " Continue to use the Tosoh AIA assays listed in notice only for patients who have not undergone a fluorescein angiography procedure exam, or on patient samples which have been taken prior to receiving the fluorescein angiography procedure. " Complete & return the Urgent Medical Device Recall Acknowledgement Form within 15-days of receiving the notification. " Maintain a copy of the notification with your laboratory records and forward this information to others who may have received this product. Should you have any questions, please contact Bernadette OConnell at 1(800) 248-6764 or by email at Bernadette.oconnell@tosoh.com. She will be able to answer any questions Monday-Friday from 9 AM to 5 PM (PST).
Quantity
992 units
Distribution
US Nationwide
Lot, serial or model codes
all lots
510(k) numbers
["K121533"]
PMA numbers
not on file
Product code
LPS
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register