Skip to the content

FDA Medical Device Recalls (openFDA)

Rymed Technologies, LLC: InVision-Plus and InVision-Plus Junior Needleless IV Connector, Disposable, Single Use Products are intended for single patient use in intravenous and blood administration sets without the need for needles, thus eliminating the potential for needle-stick injuries during use.

Recall number
Z-0675-2016
Recalling firm
Rymed Technologies, LLC
Product
InVision-Plus and InVision-Plus Junior Needleless IV Connector, Disposable, Single Use Products are intended for single patient use in intravenous and blood administration sets without the need for needles, thus eliminating the potential for needle-stick injuries during use.
Status
Terminated
Initiated
2016-01-05
Posted
2016-01-21
Terminated
2016-10-19
Reason
Potential for the female luer component of the InVision-Plus IV connector to crack during use.
Root cause
Process design
Action
The firm Ry Med Technology LLC, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 4, 2016 and response form to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use and return the product to: RyMed Technologies LLC, 6000 W. William Cannon Drive, Building B, Suite 300 Austin, TX 78749; the user facilities to change the IV connectors in 24 hour when lipids are being administered; if further distributed, contact other accounts, advise of recall and return outstanding stock; complete and return the RECALL NOTIFICATION ASAP and no later than 10 business day via email to: amccutchen@rymedtech.com, fax to: 512-301-7338, Attention Recall Return Notification or mail to RyMed Technologies LLC, Attention Recall Return Notification, 6000 W. William Cannon, Building B, Suite 300 Austin, TX 78749. If you have any questions concerning the recall, contact Director of Quality Assurance/Regulatory Affairs by email to: amccutchen@rymedtech.com or call 512-301-7338x303. If you have questions about returning/replacement of product contact Logistics Specialist by email to: svara@rymedtech.com or call 512-301-7338x309.
Quantity
496,400 units
Distribution
US Nationwide Distribution.
Lot, serial or model codes
InVision-Plus Catalog number RYM-5000 Lots 610, 611, 613, 614; Catalog number RYM-5001 Lots R01777, R01789, R01807, R01817, R01822, R01827, R01839, R01848, R01857, R01860, R01871, and R01874; Catalog number RYM-5001CL Lots R01825, R01887, and R01890. InVision Plus Junior Catalog number RYM-8001 Lot R01853.
510(k) numbers
["K991653"]
PMA numbers
not on file
Product code
FPA
Firm FEI
3005951712
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register