Philips Medical Systems (Cleveland) Inc: Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.

Recall number
Z-0675-2013
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.
Status
Terminated
Initiated
2012-09-11
Posted
2013-01-16
Terminated
2014-02-14
Reason
Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.
Root cause
Software design
Action
Philips Healthcare sent a "CUSTOMER INFORMATION -URGENT-MEDICAL DEVICE CORRECTION" letter dated September 13, 2012. The letter identified the product, problem, and actions to be taken by the customers. Philips Service Engineers will contact customers for implementation of the software update. Contact your local Philips representative or local Philips office for further information concerning this notice.
Quantity
108 units
Distribution
Worldwide Distribution-USA including the states of AZ, CA, CO, IL, IL, IN, MA, ME, MI, MO, NC, NJ, NY, OR, TN, TX, VA and WI and the countries of Austria, Canada, China, France, India, Ireland, Italy, Japan, Netherlands, Norway, Spain, Switzerland, Thailand and United Kingdom.
Lot, serial or model codes
Model #728269; Serial #s: 79906, 79921, 79927, 79937, 80007, 80026, 80031, 80040, 80058, 80062, 80066, 80069, 80087, 80110, 80121, 80155, 80167, 80168, 80169, 80174, 80192, 80193, 80195, 80201, 80208, 80209, 80219, 80227, 80228, 80230, 80252, 80258, 80266, 80280, 80289, 80293, 80294, 80315, 80322, 80328, 80333, 80340, 80341, 80348, 80360, 80375, 80385, 80387, 80394, 80398, 80416, 80421, 80422, 80429, 80447, 80451, 80460, 80474, 80482, 80484, 80485, 80492, 80497, 80505, 80516, 80578, 80585, 80591, 80629, 80646, 80658, 80681, 80704, 80709, 80711, 80720, 80723, 80743, 80749, 80751, 80831, 80835, 80866, 80878, 80908, 80909, 80911, 80914, 80916, 80930, 80931, 80944, 80974, 80992, 80999, 81023, 81025, 81034, 81045, 81071, 81099, 81101, 81107, 81117, 81126, 81136, 95382 & 95592.
510(k) numbers
["K961464"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register