Skip to the content

FDA Medical Device Recalls (openFDA)

American Medical Systems: her option Cryoablation Therapy Disposable Probe, Model CU-1. Sterile. Manufactured at: American Medical Systems, Inc., 10700 Bren Road West, Minnetonka, MN 55343 USA. The Her Option Cryoablation Therapy System consists of a Console, a Cryoprobe and a Disposable Probe.

Recall number
Z-0675-06
Recalling firm
American Medical Systems
Product
her option Cryoablation Therapy Disposable Probe, Model CU-1. Sterile. Manufactured at: American Medical Systems, Inc., 10700 Bren Road West, Minnetonka, MN 55343 USA. The Her Option Cryoablation Therapy System consists of a Console, a Cryoprobe and a Disposable Probe.
Status
Terminated
Initiated
2006-01-13
Posted
2006-03-23
Terminated
2008-02-24
Reason
Overheating of the Cryoablation Therapy Disposable Probe (CU-1), could cause a burn injury to a patient or user in certain circumstances.
Root cause
Other
Action
An Urgent Device Recall / Product Safety Advisory letter, dated January 25, 2006, was sent to all customers who have purchased Disposable Probes since 01/01/2005 or who have in their possession a Her Option console (whether customer owned or AMS owned). The letter informs customers of the issue and of the 2 lots being recalled. The letter also includes Safety Advisory Instruction which are to be posted on or near the Her Option Console. A Recall Acknowledgement Form is asked to be returned to American Medical Systems which serves as an indication that the customer read the notice, is returning affected product and have understood and posted the Safety Advisory instructions.
Quantity
152
Distribution
Nationwide. AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI
Lot, serial or model codes
Lots 431803 & 436666
510(k) numbers
not on file
PMA numbers
["P000032"]
Product code
MNB
Firm FEI
1000116179
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register