Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: Convertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL

Recall number
Z-0674-2025
Recalling firm
Cardinal Health 200, LLC
Product
Convertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL
Status
Open, Classified
Initiated
2024-11-01
Posted
2024-12-11
Terminated
not on file
Reason
Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
Root cause
Nonconforming Material/Component
Action
An URGENT MEDICAL DEVICE PRODUCT RECALL notification letter dated 11/1/24 was sent to customer. Actions Required: 1. REVIEW your inventory for the affected product codes. Location of product code and is shown on the label below (Attachment 1). 2. COMMUNICATE with all personnel that utilize the above listed Cardinal Health AAMI Level 3 Isolation Gowns. 3. SEGREGATE and quarantine all affected product upon review of your inventory. Affected product should not be used. Utilize return directions below to return product. 4. DISSEMINATE this notice to all departments, clinics and external campuses that handle the affected products. 5. DISTRIBUTORS please notify any customers to whom you may have distributed/ forwarded affected product to or will send the product on to about this medical device product recall and share a copy of this notice. 6. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email directly to GMB-FieldCorrectiveAction@cardinalhealth.com, whether you have affected product or not. Return of Product and Available Assistance: CONTACT the appropriate Customer Service group with questions relating to credit, replacement, and suitable alternative products. Monday Friday between 8:00am - 5pm EST: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All Other Customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Quantity
940,203 total units
Distribution
US Nationwide. Canada, UAE
Lot, serial or model codes
All Lots. AT6100 UDI-DI: CS (01) 50192253048252 EA (01) 20192253048251; AT6100-XL UDI-DI: CS (01) 50192253048269 EA (01) 20192253048268; AT6200 UDI-DI: CS (01) 50192253048276 EA (01) 20192253048275; AT6200-XL UDI-DI: CS (01) 50192253048283 EA (01) 20192253048282
510(k) numbers
["K160339"]
PMA numbers
not on file
Product code
FYC
Firm FEI
3001236905
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register