Beckman Coulter Inc.: Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.

Recall number
Z-0673-2025
Recalling firm
Beckman Coulter Inc.
Product
Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
Status
Open, Classified
Initiated
2024-11-08
Posted
2024-12-11
Terminated
not on file
Reason
The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
Root cause
Process design
Action
On November 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE RECALL letters that the affected assay may produce falsely decreased results if EDTA and lithium heparin plasma samples are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the test. Access Intact PTH assay Routine Mode and Intraoperative Mode users can select any of the following options: . Use serum as the sample type. . Perform testing on an alternate methodology, such as the Access 2 or UniCel DxI 600 or 800 Immunoassay Analyzers. . Continue to use EDTA and/or lithium heparin plasma sample types if the tests are completed within the timeframes listed below upon sample presentation to the DxI 9000 Access Immunoassay Analyzer (see page 3 for instructions): o Access Intact PTH assay Routine Mode: 25 minutes. o Access Intact PTH assay Intraoperative Mode: 20 minutes. . Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to perform a retrospective review of the patient results generated from Access Intact PTH Routine Mode plasma samples on the DxI 9000 to determine if retesting is necessary.
Quantity
31 US; 167 OUS
Distribution
Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, S. Korea, Spain, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.
Lot, serial or model codes
UDI-DI 15099590201937 Lots 338921 339071 372223 439138 439205 439206 439207 439650 439801 439889 440026 440090 472012 472092 472136 472154
510(k) numbers
["K232791"]
PMA numbers
not on file
Product code
CEW
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register