Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker GMRS Proximal Femoral Surgical Protocol; Global Modular Replacement System Proximal Femoral Resection for Large Segmental Replacements, Femur Prosthesis Surgical Literature, Stryker Orthopaedics, Mahwah, NJ 07430

Recall number
Z-0668-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker GMRS Proximal Femoral Surgical Protocol; Global Modular Replacement System Proximal Femoral Resection for Large Segmental Replacements, Femur Prosthesis Surgical Literature, Stryker Orthopaedics, Mahwah, NJ 07430
Status
Terminated
Initiated
2007-12-18
Posted
2008-04-22
Terminated
2008-10-15
Reason
The GMRS Proximal Femoral Surgical Protocol states " The GMRS Proximal Femoral components are fully compatabile with all Stryker V40 femoral heads." This statement is incorrect and conflicts with the Alumina Head label and packaging insert.
Root cause
Labeling design
Action
Stryker sent Urgent Product Recall notification letters and Product Accountability forms via FedEx on 12/21/2007, return receipt requested. An Urgent Product Recall letter dated 1/7/2008, was sent to surgeons who are GMRS users, advising that the GMRS Proximal femoral components are not compatible with Alumina ceramic heads.
Quantity
5869
Distribution
Worldwide Distribution - USA, Canada, Mexico, China, Australia, Hong Kong, Singapore, Sweden, Germany, Brussels, the Netherlands, Italy, Poland, Dubai, United Kingdom, Spain, and South Africa.
Lot, serial or model codes
Literature number: LSPK41; 10 surgical protocols per package.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JDI
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register