Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 (with 4 inch diameter roller pump); Catalog number 801040.

Recall number
Z-0666-05
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Advanced Perfusion System 1 (with 4 inch diameter roller pump); Catalog number 801040.
Status
Terminated
Initiated
2005-03-17
Posted
2005-03-31
Terminated
2005-09-16
Reason
The pumps will stop, show an overspeed error message, and will not restart using routine procedures if the user reduces the pump speed to zero RPM while the pump is operating in pulse mode.
Root cause
Other
Action
Recall letters dated 3/17/05 were sent to customers advising them of the problem, the temporary fix, and that software will be upgraded.
Quantity
not on file
Distribution
Nationwide, Australia, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malasia, Mexico, Pakistan, South Africa, South Korea, Taiwan, Thailand and United Arab Emirates.
Lot, serial or model codes
Serial numbers 0028 through 0527 and 0700 through 0874.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register