Becton Dickinson & Company: BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Recall number
Z-0665-2022
Recalling firm
Becton Dickinson & Company
Product
BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200
Status
Open, Classified
Initiated
2022-01-06
Posted
not on file
Terminated
not on file
Reason
May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts
Root cause
Under Investigation by firm
Action
BD issued Urgent Medical Device Correction letter via email to Sysmex on January 06, 2022, requesting they inform/distribute to their customers still using a Sysmex UF-1000i or UX-2000 analyzer the notification provided by BD .The scope would include any customers who were shipped Vacutainer LLAD products still using a Sysmex UF-1000i or UX-2000 analyzer. Sysmex issued letter to their customers addressed to the Urinalysis Manager instructs the customer to respond to Sysmex. 1. Sysmex customers should retain a copy of this notification and post near your analyzer as a visual reminder of this potential issue. 2. Complete the Notification Response Form to confirm receipt of this medical device correction. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program. 1. BD is updating the IFU for the LLAD products to include a precaution to confirm any abnormal urinalysis results per facility policy/procedure and best practices and that BD Vacutainer LLAD devices are not recommended for use with Sysmex UF-1000i and UX- 2000 analyzers as it may falsely identify contaminant particles as Red Blood Cells. BD will notify Sysmex via follow-up communication when these revisions are completed (anticipated by June 2022). 2. BD continues to investigate the root cause of this issue and will apply corrective actions as appropriate. Sysmex will be notified if any additional actions are required upon completion of the root cause analysis. Product will continue to ship so as not to impact critical patient care.
Quantity
288,152,790 Units
Distribution
US Nationwide distribution in the state of IL.
Lot, serial or model codes
All Lots within Expiry Catalog No: 364902 UDI: (01)50382903649025 Catalog Number: 36490200 UDI:(01)60382903649022
510(k) numbers
["K991088"]
PMA numbers
not on file
Product code
JKA
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register