Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular, LLC: CARDIOSAVE Intra-Aortic Balloon Pump An electromechanical system used to inflate and deflate intra-aortic balloons. It provide temporary support to the left ventricle via the principle of counterpulsation.

Recall number
Z-0662-2013
Recalling firm
Maquet Cardiovascular, LLC
Product
CARDIOSAVE Intra-Aortic Balloon Pump An electromechanical system used to inflate and deflate intra-aortic balloons. It provide temporary support to the left ventricle via the principle of counterpulsation.
Status
Terminated
Initiated
2012-03-27
Posted
2013-01-09
Terminated
2015-09-30
Reason
In the Cardiosave Intra-aortic Balloon Pump, there is a possibility that the coiled cord assembly, a prominent connector cord between the Cardiosave IABP housing and the dual display head which secured atop the IABP may fail as a result of mishandling.
Root cause
Device Design
Action
Maquet Cardiovascular sent an Urgent Product Recall Letter dated March 27, 2012, and Urgent Product Recall Acknowledgement to all affected customers . The letter identified the product, the problem, and the action to be taken by the customer. Datascope Corp determined that the best course of action was to replace the affected product. Customers were instructed that they would be contacted by a service representative to make arrangements for the removal and replacement of their affected product. Customers with questions were instructed to contact the US Call Center at 1-800-777-4222. For questions regarding this recall call 973-709-7498.
Quantity
29 units (total)
Distribution
Worldwide Distribution - USA including ID and Fl and Internationally to United Arab Emirates, France, Luxembourg, Poland, Spain, Sweden, Germany, Italy, Netherlands, Republic of China, Spain, Sweden, and Thailand
Lot, serial or model codes
510K K112372 Device Listing Number D132687   US Customer IABPs CA202159K1 CA202167K1 CA203492L1 CA203494L1  International Customer IABPs CA203486L1 CA202155K1 CA202161K1 CA203487L1 CA203490L1 CA203507L1 CA204330L1 CA204328L1 CA203505L1  International Company Owned IABPs  CA204327L1 CA202150K1 CA203506L1 CA204324L1 CA204326L1 CA202163K1 CA202169K! CA203497L1 CA203498L1 CA203496L1 CA203503L1 CA203488L1 CA203493L1 CA203508L1 CA204337L1 CA203502L1
510(k) numbers
["K112372"]
PMA numbers
not on file
Product code
DSP
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register