Skip to the content

FDA Medical Device Recalls (openFDA)

Spectranetics Corporation: Turbo Elite Atherectomy Catheter. Models: 410-152, 410-154, 414-151, 414-159, 417-152, 417-156, 420-006, 420-159, 423-001, 423-135-01 The Laser Catheters are used in conjunction with the Spectranetics CVX-300 Excimer Laser System and are intended for use in patients with single or multivessel coronary artery disease, either as a stand-alone modality or in conjunction with Percutaneous Transluminal Coronary Balloon Angioplasty (PTCA), and who are acceptable candidates for coronary artery bypass graft (CABG) surgery.

Recall number
Z-0661-2017
Recalling firm
Spectranetics Corporation
Product
Turbo Elite Atherectomy Catheter. Models: 410-152, 410-154, 414-151, 414-159, 417-152, 417-156, 420-006, 420-159, 423-001, 423-135-01 The Laser Catheters are used in conjunction with the Spectranetics CVX-300 Excimer Laser System and are intended for use in patients with single or multivessel coronary artery disease, either as a stand-alone modality or in conjunction with Percutaneous Transluminal Coronary Balloon Angioplasty (PTCA), and who are acceptable candidates for coronary artery bypass graft (CABG) surgery.
Status
Terminated
Initiated
2016-10-31
Posted
2016-11-28
Terminated
2017-04-18
Reason
Spectranetics Corporation announces a voluntary field action for the Turbo Elite Atherectomy Catheter, ELCA Coronary Atherectomy Catheter, and Bridge Occlusion Balloon due to potentially compromised integrity of the outer sterile packaging.
Root cause
Packaging
Action
Spectranectics sent an Urgent Medical Device Recall letter dated October 31, 2016, to all affected consignees. Consignees were instructed to discontinue use of the potentially impacted lots and return the product for replacement. Consignees were also instructed to complete the attached Acknowledgement and Receipt Form to facilitate the product return of any remaining product in inventory. Customers with questions were instructed to call 1-800-231-0978, Option 2. For questions regarding this recall call 719-651-8517.
Quantity
14,624 units
Distribution
Worldwide Distribution - US (nationwide and PR) Internationally to Austria, Belgium, Cyprus, Denmark, France, Germany, Hungary, Italy, Netherlands, Norway, Poland, Spain, Sweden, Switzerland, United Kingdom, Australia, Bahrain, Brazil, Canada, Chile, Colombia, Dominican Republic, Israel, Japan, Kuwait, Taiwan, Thailand.
Lot, serial or model codes
FBD16D01A; FBD16F20A; FBD16D11A; FBD16F29A; FBD16D14A; FBD16G18A; FBD16D23A; FBD16G29A; FBD16E05A; FBD16H03A; FBD16E11A; FBD16H26A FBD16E19A; FBD16J06A; FBD16E26A; FBD16J21A; FBD16F03A; FBD16J26A; FBE16C30A; FBE16H24A; FBE16E25A; FBE16J14A; FBE16F15A; FAY16C29A; FAY16G14A; FAY16D08A; FAY16G29A; FAY16D26A; FAY16H23A; FAY16E17A; FAY16H31A; FAY16F08A; FAY16J23B; FAY16G01A; FBF16D27A; FBF16G07A FBF16E12A ; FBF16H10B; FBF16F17A; FBF16J16A; FAZ16C31A; FAZ16G25A; FAZ16D14A ; FAZ16H01A; FAZ16D25A; FAZ16H22A; FAZ16E06A; FAZ16H30A FAZ16F24B; FAZ16J20A; FAZ16F27A; FBG16E02A; FBG16H31A; FBG16F23A; CBA16H19A; FBA16F27A; CBA16H19B; FBA16G11A; CBA16H19C; FBA16G15A; FBA16D04A; FBA16G19A; FBA16D06A; FBA16G27B; FBA16D18A; FBA16H02A FBA16D21A; FBA16H08A; FBA16E09A; FBA16H17A; FBA16E10A; FBA16J02A FBA16E13A; FBA16J07A; FBA16E19A; FBA16J19A; FBA16F07A; FBA16J22A; FBA16F21A; FBE16E02A; FBE16G06A; FBE16G20A; FAZ16E26A; FAZ16G12A; FBA16D29A; FBA16F16B;
510(k) numbers
["K140775"]
PMA numbers
not on file
Product code
MCW
Firm FEI
3007284006
Firm city
Colorado Springs
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register