Skip to the content

FDA Medical Device Recalls (openFDA)

Fukuda Denshi USA, Inc.: Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is meant to acquire and monitor physiological signals from patients.

Recall number
Z-0661-2016
Recalling firm
Fukuda Denshi USA, Inc.
Product
Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is meant to acquire and monitor physiological signals from patients.
Status
Terminated
Initiated
2015-12-21
Posted
2016-01-16
Terminated
2016-08-19
Reason
A possible malfunction where a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term Backup battery).
Root cause
Process control
Action
Fukuda Denshi USA, Inc. of Redmond, WA plans to perform 100% field correction of units distributed in the US. The Recall/Field Correction Notice letter (FCN-A01-201512), dated December 21, 2015, and the Customer Confirmation Sheet will be sent to consignees using a clearly red-marked URGENT via Express Mail. The consignees are requested to fill out the Customer Confirmation Sheet and return via fax to 888-224-7090, email to DBlakely@fukuda.com, or send via postal service back to Fukuda Denshi USA, Inc. in a pre-stamped envelope. Consignee can call Doug Blakey directly at 210-240-4521 with any questions or concerns.
Quantity
95 units distributed in the US
Distribution
Distributed in the states of CO, ID, KS, MI, MO, OR, PA, TX, UT, and WI.
Lot, serial or model codes
50400001, 50400002, 50400003, 50400004, 50400011, 50400012, 50400018, 50400019, 50400020, 50400021, 50400022, 50400023, 50400024, 50400025, 50400026, 50400027, 50400028, 50400029, 50400030, 50400031, 50400032, 50400033, 50400034, 50400035, 50400036, 50400037, 50400038, 50400039, 50400040, 50400042, 50400043, 50400044, 50400045, 50400046, 50400047, 50400048, 50400049, 50400050, 50400051, 50400052, 50400053, 50400054, 50400055, 50400056, 50400057, 50400058, 50400059, 50400060, 50400061, 50400062, 50400063, 50400064, 50400065, 50400066, 50400067, 50400068, 50400069, 50400070, 50400071, 50400072, 50400073, 50400074, 50400075, 50400076, 50400077, 50400078, 50400084, 50400093, 50400094, 50400095, 50400096, 50400097, 50400098, 50400099, 50400100, 50400101, 50400102, 50400103, 50400104, 50400105, 50400106, 50400107, 50400108, 50400109, 50400110, 50400111, 50400112, 50400113, 50400114, 50400115, 50400116, 50400117, 50400118, 50400119, and 50400120.
510(k) numbers
["K134046"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3000215024
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register