Skip to the content

FDA Medical Device Recalls (openFDA)

CELLTRION USA INC: Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

Recall number
Z-0659-2022
Recalling firm
CELLTRION USA INC
Product
Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02
Status
Terminated
Initiated
2021-12-28
Posted
2022-03-02
Terminated
2024-06-10
Reason
The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021.
Root cause
Process control
Action
The firm initially placed a hold on product due to complaints of false positive and false negative results in September 2021. After product was erroneously distributed from this lot in December 2021, the firm notified affected customers that the lot should not have been distributed and should not be used. This email went out on or around December 28, 2021. Customers were instructed to NOT use the affected lot.
Quantity
243 kits (6,075 tests)
Distribution
US distribution to CA, FL, NJ, NY, PA, SC.
Lot, serial or model codes
Lot COVGCCM0008
510(k) numbers
not on file
PMA numbers
not on file
Product code
QKP
Firm FEI
3015423092
Firm city
Jersey City
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register