Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Biopsy Needle FPU Kit, Product # 9736075, UDI: 00763000063108

Recall number
Z-0659-2020
Recalling firm
Medtronic Navigation, Inc.
Product
Biopsy Needle FPU Kit, Product # 9736075, UDI: 00763000063108
Status
Open, Classified
Initiated
2019-10-30
Posted
not on file
Terminated
not on file
Reason
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
Root cause
Process control
Action
On 10/24/2019, Medtronic sent a "URGENT: MEDICAL DEVICE RECALL" notification to consignees via FedEx. The customer notification letter as consignees to do the following: 1. Identify, segregate, and quarantine affected products within your inventory 2. Complete the Customer Confirmation Form. Return the form to Medtronic via email at RS.NavFCA@Medtronic.com or via fax at 651-367-7075 within 30 days of receipt. 3. Contact Medtronic at 1-888-826-5603 to receive a return materials authorization (RMA) and schedule replacement product. Return affected products to: 4. Medtronic Product Services, Attention: RMA #, 1480 Arthur Ave Louisville, CO 80027 5. If you have questions E-mail Medtronic at RS.NavTechSupport@Medtronic.com or call 1-888-826-5603.
Quantity
1 disposable Biopsy Needle
Distribution
US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jersey,New Mexico,New York,North Carolina,North Dakota,Ohio,Oklahoma,Oregon,Pennsylvania,South Carolina,South Dakota,Tennessee,Texas,Utah,Vermont,Virginia,Washington,West Virginia,Wisconsin,Wyoming, OUS: None: Australia, Austria,Bangladesh,Belgium,Bosnia And Herzegovina,Brazil,Bulgaria,Canada,Canary Islands,China,Colombia,Costa Rica,Croatia,Czech Republic,Denmark,Ecuador,Egypt,Finland,France,Germany,Greece,Hungary,Iceland,India,Indonesia,Ireland,Israel,Italy,Japan,Jordan,Kazakhstan,Kenya,Korea, Republic Of,Latvia,Lebanon,Luxembourg,Malaysia,Martinique,Mexico,Netherlands,New Zealand,Pakistan,Philippines,Poland,Portugal,Puerto Rico,Qatar,Reunion,Romania,Russian Federation,Saudi Arabia,Serbia,Slovakia,Slovenia,South Africa,Spain,Sweden,Switzerland,Thailand,Turkey,United Arab Emirates,United Kingdom
Lot, serial or model codes
Lot Numbers: 0009754921 0009921274 0009940846 0009921273 0009921276 0009952564
510(k) numbers
["K971247"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register