Stryker Communications, Inc.: (l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the Stryker Spring Arm used in multiple configurations as a component of the (1)Visum Halogen Surgical Light, (2)Visum LED Surgical Light, and (3)Stryker Flat Panel and Navigation Arm. The Spring Arm that is subject to this field action suspends loads, is a conduit for power/data, and facilitates multi-directional movement. The Spring Arms that have been identified within scope are replacement parts that have potentially been serviced incorrectly.

Recall number
Z-0657-2013
Recalling firm
Stryker Communications, Inc.
Product
(l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the Stryker Spring Arm used in multiple configurations as a component of the (1)Visum Halogen Surgical Light, (2)Visum LED Surgical Light, and (3)Stryker Flat Panel and Navigation Arm. The Spring Arm that is subject to this field action suspends loads, is a conduit for power/data, and facilitates multi-directional movement. The Spring Arms that have been identified within scope are replacement parts that have potentially been serviced incorrectly.
Status
Terminated
Initiated
2012-09-25
Posted
2013-01-09
Terminated
2013-04-22
Reason
Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
Root cause
Equipment maintenance
Action
Stryker sent an Urgent Medical Device Correction letter dated September 25, 2012, to all affected customers. The letter identified the product, the prolem, and the action to be taken by the customer. Customers were instructed to immediately inspect their Spring Arm. If customers experienced any concerns with their Spring Arm they were instructed to contact Stryker immediately at 1-866-841-5663 to schedule a service technician to inspect and correct the Spring Arm circlips. For questions regarding this recall call 972-410-7100.
Quantity
United States: 1,082 units and Outside United States: 199 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.
Lot, serial or model codes
Code Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG,  0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSY
Firm FEI
3002975342
Firm city
Flower Mound
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register