Skip to the content

FDA Medical Device Recalls (openFDA)

Nova Ortho-Med Inc: Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)

Recall number
Z-0654-2021
Recalling firm
Nova Ortho-Med Inc
Product
Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)
Status
Open, Classified
Initiated
2019-10-07
Posted
not on file
Terminated
not on file
Reason
Walkers within the identified serial range have the potential for the front fork to unscrew itself from the leg. A partially unscrewed fork creates unevenness in the wheelbase, which could cause instability, that if not detected by the user, may result in the front fork falling off the walker.
Root cause
Device Design
Action
On 10/07/2019, Product Correction and Product Safety Recall notices were emailed to customers. Customers were asked to replace the front leg assemblies as follows: a. Depress the push buttons on the original legs and remove them from the frame. b. Insert the improved replacement assemblies issued by NOVA into the unit and discard the original legs. In addition, customers were asked to do the following: 1. Check your inventory for affected walkers and fix. 2. Contact all customers who purchased affected walkers notify them of the potential issue. Schedule a time that they can come in for repair or if the customer prefers the recalling firm can send the wheel assemblies directly to their residence. Have them inspect their walker immediately to identify if either of the front forks have unscrewed from the leg. If so, please instruct them to not use the walker until the leg assemblies can be replaced with the new parts. 3. Call the recalling firm to request the wheel assembly set to be sent to the required location. 4. Record the serial #'s of the product that have been corrected and email them to Customer Service (info@novajoy.com) so that we can have for our records. If you prefer to call, then dial 1-800-557-6682 x8129. Email: info@novajoy.com
Quantity
10898
Distribution
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, WY O.U.S.: Taiwan
Lot, serial or model codes
Serial # range: HF-4258BKHA0001 to HF4258RDHK0352 (Purchased before April 4th,2019)
510(k) numbers
not on file
PMA numbers
not on file
Product code
ITJ
Firm FEI
3002329391
Firm city
Carson
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register