Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Zimmer Universal Locking System 3.5mm Locking Reconstruction Plate Straight 11 Holes 144mm Length Zimmer Universal Locking System 3.5mm Locking Dual Compression Plate 8 Holes 105mm Length. Indicated for temporary internal fixation and stabilization of osteotomies and fractures.

Recall number
Z-0654-2016
Recalling firm
Zimmer Biomet, Inc.
Product
Zimmer Universal Locking System 3.5mm Locking Reconstruction Plate Straight 11 Holes 144mm Length Zimmer Universal Locking System 3.5mm Locking Dual Compression Plate 8 Holes 105mm Length. Indicated for temporary internal fixation and stabilization of osteotomies and fractures.
Status
Terminated
Initiated
2015-12-09
Posted
not on file
Terminated
2016-11-08
Reason
The product on the package label did not match what was inside the package. The label was for part number 00-4936-011-13 lot number 62968956 while the product inside the package was part number 00-4936-008-07 lot number 62968974.
Root cause
Labeling Change Control
Action
On 12/9/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital risk managers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Specifically, customers were asked to review the notification and ensure all affected personnel are aware of the information; assist the Zimmer Biomet sales representative with the quarantine and removal of any of the identified affected product; and complete and return the Certificate of Acknowledgement form. If after reviewing this notification customers have further questions or concerns please call the customer call center at 1-877-946-2761. Hours of operation are Monday through Friday, 8 a.m. through 5 p.m. EST.
Quantity
52
Distribution
Worldwide Distribution -- USA, including the states of MN, NY, MS, AL, OH, KY, ME, and NM; and, the countries of Taiwan, Korea, Singapore, El Salvador, Uruguay, Czech Republic, France, Italy, Poland, United Kingdom, South Africa, and Slovakia.
Lot, serial or model codes
Part 00-4936-011-13, Lot 62968956 3.5MM ULS RECON PLATE 11H Part 00-4936-008-07, Lot 62968974 3.5MM ULS DUAL COMP PLT 8
510(k) numbers
["K060710"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register