FDA Medical Device Recalls (openFDA)
Surgical Tissue Network, Inc.: UltraFill DBM labeled in part: TISSUENET Sterilized by T10^6 Description: UltraFill DBM Size: 5cc Product Code: RT53005 Donated Human Tissue Allograft SINGLE PATIENT USE ONLY FOR INTERNATIONAL USE ONLY See Package Insert for Additional Information TissueNet's Porcine DBM product line is used as a bone void filler.
- Recall number
- Z-0652-2013
- Recalling firm
- Surgical Tissue Network, Inc.
- Product
- UltraFill DBM labeled in part: TISSUENET Sterilized by T10^6 Description: UltraFill DBM Size: 5cc Product Code: RT53005 Donated Human Tissue Allograft SINGLE PATIENT USE ONLY FOR INTERNATIONAL USE ONLY See Package Insert for Additional Information TissueNet's Porcine DBM product line is used as a bone void filler.
- Status
- Terminated
- Initiated
- 2012-03-14
- Posted
- 2013-01-07
- Terminated
- 2014-12-02
- Reason
- Surgical Tissue Network Inc., DBA TissueNet Inc. recalled their UltraFill DBM Putty-Porcine (1 cc; 5cc; 10cc) because products may contain trace amounts of 316L Stainless Steel particulates.
- Root cause
- Employee error
- Action
- Surgical Tissue Network sent a Notification of Voluntary Tissue Recall dated March 22, 2012, to all affected customers. The firm issued an additional notification letter on August 30, 2012. The letter identified the product, the problem, and action to be taken by the customer. Consignees were asked to return all listed products that remain in inventory. If the product was further distributed, they were asked to forward the recall information to their customers. Customers were instructed to contact TissueNet's Customer Service Department at 800-465-8800 x283 to coordinate return/replacement of the affected product. Customers with questions were instructed to call 800-465-8800 x230. For questions regarding this recall call 407-380-2424.
- Quantity
- 208 units
- Distribution
- Worldwide Distribution including Turkey, Greece, & Mexico.
- Lot, serial or model codes
- Product Code: RT53005 Tissue ID: TN48005-11-0021 through TN48055-11-0059; TN48055-11-0070 through TN48055-11-0075; TN58063-11-0101 through TN58063-11-0142; TN61039-11-0031 through TN61039-11-0097; TN70037-11-0017 through TN70037-11-0018; TN70037-11-0020 through TN70037-11-0037; TN74060-11-0021 through TN74060-11-0047; TN74060-11-0050 through TN74060-11-0056
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MQV
- Firm FEI
- 3003564213
- Firm city
- Orlando
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28