Skip to the content

FDA Medical Device Recalls (openFDA)

California Medical Laboratories Inc: Rigid Tip Suction Wand Non sterile bulk devices (S112468N, S120196N, and S120197N) were sold to Terumo UK to be included in a kit under Terumo as manufacturer's name. The Suction Wands are indicated for use to remove excess fluid from the surgical field.

Recall number
Z-0650-2013
Recalling firm
California Medical Laboratories Inc
Product
Rigid Tip Suction Wand Non sterile bulk devices (S112468N, S120196N, and S120197N) were sold to Terumo UK to be included in a kit under Terumo as manufacturer's name. The Suction Wands are indicated for use to remove excess fluid from the surgical field.
Status
Terminated
Initiated
2012-11-02
Posted
2013-01-07
Terminated
2013-03-04
Reason
California Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the potential to come loose from the wand and fall into the patient's chest cavity.
Root cause
Nonconforming Material/Component
Action
CalMed sent an Urgent Field Safety Notice dated November 2, 2012, to all affected customers by certified mail and additionally followed up via electronic mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact Sorin Group USA's Customers Service Department at +1-800-650-2623 to return any unused product and order replacements. Customers were also asked to communicate the important safety information to all potential users within their organization or any other organization where the affected product may have been transferred and to maintain awareness of this communication for an appropriate period of time to ensure effectiveness of the corrective action. Customers were asked to return all unused product to CalMed Labs. Customers with questions were instructed to contact California Medical Laboratories at +1-714-556-7365. For questions regarding this recall call 714-556-7365.
Quantity
S112468N=1000 units, S120196N=1400 units, & S120197N=1200 units
Distribution
Worldwide Distribution - USA including AZ, FL, CA, LA, CO, and MS. Internationally to Canada, Israel, Greece, Hong Kong, India, Ireland, Jordan, Lebanon, Malaysia, Netherlands, Russia, Saudi Arabia, South Africa and Spain.
Lot, serial or model codes
Catalog No: SU-12202NS Lot No: S112468N, S120196N, S120197N
510(k) numbers
["K982891"]
PMA numbers
not on file
Product code
DTS
Firm FEI
1000135553
Firm city
Costa Mesa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register