Skip to the content

FDA Medical Device Recalls (openFDA)

Optimedica Corporation: Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

Recall number
Z-0649-2014
Recalling firm
Optimedica Corporation
Product
Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.
Status
Terminated
Initiated
2013-12-20
Posted
2014-01-08
Terminated
2014-06-26
Reason
Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.
Root cause
Under Investigation by firm
Action
The firm, OptiMedica, issued an "ADVISORY NOTICE" dated December 20, 2013 to all customers with a Catalys Precision Laser System via Federal Express. OptiMedica also released a Field Service Bulletin (Document No: SB-00065) internally for the impacted product with instructions to implement the corrections through the OptiMedica service organization. The notice describes the product, problem and actions to be taken. The customers were instructed to complete the site header information on the attached fax-back form and return via fax to 1-408-890-4681 within the next three (3) business days or mail to OptiMedica at 1310 Moffette Park Drive, Sunnyvale, CA 94089. If you have any questions related to the operation of the Catalys Precision Laser System or the documentation requirements associated with this letter, please call OptiMedica Technical Support at 1-855-662-0911 and you will be directed to an OptiMedica representative who will be pleased to assist you.
Quantity
78 (39 in US, 39 internationally)
Distribution
Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.
Lot, serial or model codes
All units shipped prior to Octotber 18, 2013. US Serial numbers: 44055311, 44055611, 44055811, 44056012, 44056312, 44056612, 44056712,  44056812, 44057312, 44057812, 44058012, 44058212, 44058312, 44059212, 44059312, 44059612, 44059712, 44059812, 44059912, 44060012, 44060112, 44060312, 44060412.  International Serial numbers: 44000811, 44055211, 44055411, 44055711, 44055911, 44056112, 44056212, 44056412, 44056512, 44056912, 44057012, 44057112, 44057212, 44057412, 44057512, 44057612, 44057712, 44057912, 44058112, 44058512, 44058612, 44058912, 44059012. 44059112 44059412 44059512 44060212 44060512 44061112 44061212 44061312 44061512 44061712 44061913 44062113 44062313 44062513 44062913 44063013 44060612 44060712 44060912 44061012 44061412 44061612 44061812 44062013 44062213 44062613 44062813 44063313 44063513 44063613 44063713 44063913.
510(k) numbers
["K121091"]
PMA numbers
not on file
Product code
OOE
Firm FEI
3005675890
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06