FDA Medical Device Recalls (openFDA)
Brainlab AG: RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms
- Recall number
- Z-0647-2016
- Recalling firm
- Brainlab AG
- Product
- RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms
- Status
- Terminated
- Initiated
- 2015-11-05
- Posted
- not on file
- Terminated
- 2017-08-21
- Reason
- Large objects with fine resolution are potentially displayed cropped when imported into Brainlab Brain Metastases 1.0.0.
- Root cause
- Software design
- Action
- Brainlab sent a FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated November 5, 2015, to all direct accounts informing the of the recall and actions to be taken. User Corrective Action: Users of Brain Metastases Version 1.0.0 or Adaptive Hybrid Surgery Analysis 1.0.0 shall adhere to the following: After import always carefully review and verify all objects for correctness and validity. For reviewing an object, select the image set which was used to outline or to modify the objects in the Data menu. Also compare the contours of an object in Brain Metastases or Adaptive Hybrid Surgery Analysis software to the contours in SmartBrush. Please note that the error will not be visible in SmartBrush. Always verify the correct shape of the objects in Brain Metastases or Adaptive Hybrid Surgery Analysis software. If possible do not use image data sets with large slice distance and/or with a low amount of slices. To retrospectively identify if cropped objects have been used for treatment planning, the object contours of a specific treatment plan can be reviewed in Brainlab Dose Review, where the error will also be visible. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact affected customers tentatively starting February 2016 to schedule the update. The request that users advise the appropriate personnel working in their department of the content of the Notification letter. If the user facilities require further clarification, thay can contact their local Brainlab Customer Support Representative. Customer Hotline: +49 89 99 15 68 44 or +1 800 597 5911 (for US customers) E-mail: support@brainlab.com (for US customers: us.support@brainlab.com) Fax: Brainlab AG: + 49 89 99 15 68 33
- Quantity
- 20 systems
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Mexico, Portugal, South Korea, Spain, Switzerland, Turkey and the United Kingdom.
- Lot, serial or model codes
- Brain Metastases version 1.0.0 Model/catalogue number Description 21702-04 ELEMENTS AUT. BRAIN METASTASES PLAN INST 21703-01 ELEMENTS ADAPT HYB SURG ANALYSIS LICENSE 21703-02 ELEMENTS ADAPT HYB SURG ANALYSIS INST 21702 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-12 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-36 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-60 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21703 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-12 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-36 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-60 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21713 ELEMENTS ADAPTIVE HYBRID SURG. ANALYSIS 21722 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-12 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-36 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-60 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21723 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-12 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-36 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-60 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21732 ELEMENTS AUTOMATIC BRAIN METS PLAN. 12M
- 510(k) numbers
- ["K142108"]
- PMA numbers
- not on file
- Product code
- MUJ
- Firm FEI
- 3002619595
- Firm city
- Feldkirchen
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28