Skip to the content

FDA Medical Device Recalls (openFDA)

Navilyst Medical, Inc: Exodus(TM) 8F (2.7 mm)/ 35 cm Standard Loop Biliary Drainage Catheter with Hydrophilic Coating, REF/ Catalog No. 10005, UPN/Product No. H965100051, STERILE --- Package/Box Units: Box of 5 eaches --- Made in USA: 2301 Centennial Blvd. Jeffersonville, IN 47130 -- Manufactured for: Navilyst Medical, Inc. 26 Forest Street Marlborough, MA 01752 --- The Exodus Drainage Catheter consists of a radiopaque polyurethane catheter. The distal end of the catheter is configured as a J-Tip or a locking pigtail and is coated with a hydrophilic coating. Catheter markings are provided in centimeters on the shaft to indicate the distance from the distal tip. The Exodus Biliary Drainage Catheter includes a radiopaque (RO) marker band located approximate 5 mm from the most proximal drainage hole to assist with catheter placement. Biliary Drainage Catheters are intended for percutaneous drainage of the biliary tree.

Recall number
Z-0647-2012
Recalling firm
Navilyst Medical, Inc
Product
Exodus(TM) 8F (2.7 mm)/ 35 cm Standard Loop Biliary Drainage Catheter with Hydrophilic Coating, REF/ Catalog No. 10005, UPN/Product No. H965100051, STERILE --- Package/Box Units: Box of 5 eaches --- Made in USA: 2301 Centennial Blvd. Jeffersonville, IN 47130 -- Manufactured for: Navilyst Medical, Inc. 26 Forest Street Marlborough, MA 01752 --- The Exodus Drainage Catheter consists of a radiopaque polyurethane catheter. The distal end of the catheter is configured as a J-Tip or a locking pigtail and is coated with a hydrophilic coating. Catheter markings are provided in centimeters on the shaft to indicate the distance from the distal tip. The Exodus Biliary Drainage Catheter includes a radiopaque (RO) marker band located approximate 5 mm from the most proximal drainage hole to assist with catheter placement. Biliary Drainage Catheters are intended for percutaneous drainage of the biliary tree.
Status
Terminated
Initiated
2011-02-04
Posted
2012-01-11
Terminated
2012-01-10
Reason
Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.
Root cause
Nonconforming Material/Component
Action
The firm, Navilyst Medical, sent an "Urgent Medical Device Recall -Immediate Action Required" letter dated February 4, 2011, via Federal Express to the domestic consignees/customers and one foreign consignee/customer. Consignees/customers were instructed to: (1.) Immediately discontinue use and remove the recalled product from inventory; (2.) Segregate this product in a secure location for return to Navilyst Medical, Inc.; (3.) Immediately forward a copy of the recall notification to ALL sites to which they have distributed affected device (4.) promptly complete and return the Reply Verification Tracking Form (even if you do not have any product to return) via fax to: QA Manager, Navilyst Medical, Inc., 1-800-782-1357; (5.) call Customer Service at 1-800-833-9973 between 8:30.am to 7:00pm EST to obtain RMA number; (6.) package and return the recalled product; and (7.) Physicians were instructed to continue to monitor their patients and evaluate per institutional protocol for biliary catheters. If you have any questions about this recall action, your local Sales Representative is available to assist you or contact QA Manager, Navilyst Medical, Inc., directly at 518-742-4571.
Quantity
not on file
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AZ, CA, CO, FL, IL, KS, MI, MO, NC, NJ, NY, OH, OK, PA, PR and country of: South Africa.
Lot, serial or model codes
Lots 1ML0052506, 1ML0062201, 1ML0080901
510(k) numbers
["K093392"]
PMA numbers
not on file
Product code
FGE
Firm FEI
3017892510
Firm city
Glens Falls
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register