FDA Medical Device Recalls (openFDA)
Oscor Inc.: Destino Twist 14F (also branded as Guidestar 14F)
- Recall number
- Z-0646-2022
- Recalling firm
- Oscor Inc.
- Product
- Destino Twist 14F (also branded as Guidestar 14F)
- Status
- Completed
- Initiated
- 2021-12-22
- Posted
- not on file
- Terminated
- not on file
- Reason
- There is potential for separation between the threaded hub and threaded cap at the proximal end of the device.
- Root cause
- Process control
- Action
- Oscor issued customer letters to impacted consignees on January 7 and 11, 2022. Distributors are asked to immediately check inventory for affected product and quarantine any recalled devices. All downstream customers (hospitals/end users) should be identified and this recall should be communicated to those customers immediately. All retrieved inventory is to be returned to Oscor. An acknowledgement form must be completed and returned to Oscor via fax (727-934-9835) or via email to recall@Oscor.com.
- Quantity
- 549 units
- Distribution
- Distribution to TN and PA in USA OUS distribution to Czech Republic
- Lot, serial or model codes
- 1) Model DST1405525 (Destino Twist 14F) - a) Lot: PQOC13475 [UDI: (01)00885672011123(11)211014(17)241001(10)PQOC13475] 2) Model D141103 (Guidestar 14F) - Lots: a) DP14073 - [UDI: (01)00885672010911(11)211114(17)241101(10)DP14073] b) DP14072 - [UDI: (01)00885672010911(11)211024(17)241001(10)DP14072] c) DP14071 - [UDI: (01)00885672010911(11)211023(17)241001(10)DP14701] d) DP13834 - [UDI: (01)00885672010911(11)210919(17)240901(10)DP13834] e) DP13828 - [UDI: (01)00885672010911(11)210914(17)240901(10)DP13828] f) DP13827 - [UDI: (01)00885672010911(11)210826(17)240801(10)DP13827] g) DP13811 - [UDI: (01)00885672010911(11)210812(17)240801(10)DP13811] h) DP13812 - [UDI: (01)00885672010911(11)210812(17)240801(10)DP13812] i) PQOC12995 - [UDI: (01)00885672010911(11)210709(17)240701(10)PQOC12995] j) PQOC12632 - [UDI: (01)00885672010911(11)210706(17)240701(10)PQOC12632] k) PQOC12633 - [UDI: (01)00885672010911(11)210705(17)240701(10)PQOC12633] l) PQOC12345 - [UDI: (01)00885672010911(11)210622(17)240601(10)PQOC12345] m) PQOC12330 - [UDI: (01)00885672010911(11)210602(17)240601(10)PQOC12330]
- 510(k) numbers
- ["K140406"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 1035166
- Firm city
- Palm Harbor
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28