FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (10 mm diameter) Model: LTF-S190-10 Used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities, including female reproductive organs.
- Recall number
- Z-0645-2024
- Recalling firm
- Olympus Corporation of the Americas
- Product
- OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (10 mm diameter) Model: LTF-S190-10 Used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities, including female reproductive organs.
- Status
- Open, Classified
- Initiated
- 2023-11-20
- Posted
- 2024-01-05
- Terminated
- not on file
- Reason
- Instructions for Use-User Deviations related to the reprocessing of the videoscope detailed in the Reprocessing Manual Instructions. The steps are required for the proper reprocessing of these videoscopes
- Root cause
- Labeling Change Control
- Action
- Olympus Medical Systems Corporation issued Urgent Medical Device Correction Letter on November 20, 2023. Letter states reason for recall, health risk and action to take: Olympus requires you to take the following actions: During an evaluation of user implementation of the Instructions for Use related to the reprocessing of the LTF-S190-5 videoscope, Olympus observed deviations from the following reprocessing steps detailed in the respective Reprocessing Manual Instructions. These steps are required for the proper reprocessing of these videoscopes. Therefore, Olympus is reminding users that following the instructions in these reprocessing manuals (LTF-S190-5 Reprocessing Manual, GE9382, and LTF-S190-10 Reprocessing Manual, RC0686) is required. Olympus is offering educational on-site support to review and train on the reprocessing instructions for the LTF-S190-5 and LTF-S190-10. The LTF-S190-10 is similar in design to the LTF-S190-5 and is therefore included in this corrective action 1. Carefully read the contents of this notification. 2. Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the reprocessing instructions contained in Sections 5.5, 5.6, and 5.7 of the LTF Reprocessing Manuals, especially adhering to the immersion time for the entire endoscope, and the brushing or wiping steps. 3. If you have further distributed this product, forward this letter to those facilities immediately. 4. Olympus requests that you acknowledge receipt of this letter. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at Olympus3074@sedgwick.com or by fax at 844-265-7383. For any questions about the acknowledgement form, please call the Sedgwick team at 844-483-3866. If you require additional information about the proper steps to reprocess, Olympus provides the following resources: Contact your local Olympus Endoscopy Support Specialist (ESS) or our Technical Assistance Center for assistance in ar
- Quantity
- 271 units US
- Distribution
- Nationwide
- Lot, serial or model codes
- UDI-DI: 04953170310386 All Serial Numbers
- 510(k) numbers
- ["K111425"]
- PMA numbers
- not on file
- Product code
- HET
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28