Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Perfusion Systems: Medtronic AFFINITY NT, Cardiotomy/Venous Reservoir (CVR) with Filter, 540 The AFFINITY(R) NT Cardiotomy/Venous Reservoir (CVR) with Filter is a single used device designed to collect and store blood during extracorporeal circulation. Venous blood is collected and de foamed while cardiotomy blood is collected, de foamed and filtered before mixing with the venous blood. The AFFINITY(R) NT Cardiotomy/Venous Reservoir (CVR) with filter is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures for up to 6 hours in duration.

Recall number
Z-0645-2018
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic AFFINITY NT, Cardiotomy/Venous Reservoir (CVR) with Filter, 540 The AFFINITY(R) NT Cardiotomy/Venous Reservoir (CVR) with Filter is a single used device designed to collect and store blood during extracorporeal circulation. Venous blood is collected and de foamed while cardiotomy blood is collected, de foamed and filtered before mixing with the venous blood. The AFFINITY(R) NT Cardiotomy/Venous Reservoir (CVR) with filter is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures for up to 6 hours in duration.
Status
Terminated
Initiated
2018-01-16
Posted
not on file
Terminated
2020-08-21
Reason
Possible sterile barrier breach in the pouch of Affinity NT Cardiotomy Venous Reservoir stand alone uncoated and Affinity NT Cardiotomy Venous Reservoir stand alone Trillium coated products.
Root cause
Process control
Action
Medtronic sent an Urgent Medical Device Recall Letter dated January 16, 2018. Customer Actions: Please review your inventory for product affected by this issue. 1.) Immediately identify and quarantine all unused, non-expired, affected product in your inventory. 2.) Return all unused, affected product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570 Option 4, or accessing My Orders and referencing this communication to initiate a return and credit of unused product. Your Medtronic sales representative can assist you in the return of affected product as necessary. 3.) Complete the enclosed Customer Confirmation Certificate and scan and email to RS.CFQFCA@medtronic.com. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. For further questions, please call (763) 526-2494.
Quantity
8644 units
Distribution
Worldwide Distribution - US Distribution and to the countries of : Australia, Canada, Colombia, Denmark, Germany, Guyana, Israel, Italy, Japan, Netherlands, Norway, Saudi Arabia and United Kingdom.
Lot, serial or model codes
All lots
510(k) numbers
["K936003"]
PMA numbers
not on file
Product code
DTN
Firm FEI
1000116158
Firm city
Brooklyn Park
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register