FDA Medical Device Recalls (openFDA)

Innovasis, Inc: Excella II Polyaxial Pedicle Screw used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:

Recall number
Z-0644-2018
Recalling firm
Innovasis, Inc
Product
Excella II Polyaxial Pedicle Screw used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:
Status
Terminated
Initiated
2018-01-08
Posted
not on file
Terminated
2018-02-27
Reason
Nine lots of 8.5 mm Polyaxial Pedicle Screws are at risk of breakage.
Root cause
Nonconforming Material/Component
Action
The distributors were notified of the recall on 1/8/2018 via email with an attached pdf. They are also being asked to confirm or deny the quantities showing in inventory via a Product Removal Form sent in the same email.
Quantity
1,010 parts
Distribution
United States
Lot, serial or model codes
Catalog/Part No.: E2S8530, E2S8535, E2S8540, E2S8545, E2S8550, E2S8555, E2S8560 with Lot Number: 1101 and Catalog/Part No.: E28540, E28545 with Lot Number: 1201. Each of these parts is engraved with part number and lot number.
510(k) numbers
["K102248"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3004719693
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register