Skip to the content

FDA Medical Device Recalls (openFDA)

CompView Medical, LLC: Model NuCARTGN. The NuCART is a mobile display and image management platform used in Minimally Invasive Surgery (MIS) environments to facilitate the routing and ergonomic display of video signals from source devices to flat-panel monitors. It is tailored for use with an OEC C-arm and Endo-Cam in MIS surgery suites, Ambulatory Surgery Centers, Urology, GI and other suites.

Recall number
Z-0644-2016
Recalling firm
CompView Medical, LLC
Product
Model NuCARTGN. The NuCART is a mobile display and image management platform used in Minimally Invasive Surgery (MIS) environments to facilitate the routing and ergonomic display of video signals from source devices to flat-panel monitors. It is tailored for use with an OEC C-arm and Endo-Cam in MIS surgery suites, Ambulatory Surgery Centers, Urology, GI and other suites.
Status
Terminated
Initiated
2015-12-04
Posted
2016-01-13
Terminated
2016-03-31
Reason
The attachment of the NuCART Spring Arm to the NuCART horizontal boom may come apart if the securing ring (snap ring) was not installed correctly causing the spring arm to release and detach.
Root cause
Nonconforming Material/Component
Action
The firm began inspecting devices between 12/4/15 and 12/16/15, and sent the Nucart Product Investigation letter to consignees between 12/11/15 and 12/15/15. Consignees with questions can call the firm at 503-601-5603.
Quantity
10 units
Distribution
US nationwide distribution including AZ, CA, CO, GA, OH, and TN.
Lot, serial or model codes
Lot# T9993-02-01 /15924CARTXXTNNA / DX00357 Lot# T9993-02-02 / 15924CARTXXAZPH / DX00358 Lot# U0328-02-02 / 15A01CARTXXCOEN /DX00360 Lot# U0328-02-01 / 15A26CARTXXCAK1 / DX00359 Lot# U0718-02-01 / 15A26CARTXXCAK2 / DX00361 Lot# U0718-02-02 / 15A26CARTXXCAK3 / DX00362 Lot# U1033-02-01 / 15A19CARTXXOHCA /DX00363 Lot# U1033-02-02 / 15A26CARTXXCASJ / DX00364 Lot# U1126-02-01 / 15A26CARTXXCAK41 /DX00365 Lot# U1126-02-02 / 15806CARTXXGAAT / DX00366
510(k) numbers
not on file
PMA numbers
not on file
Product code
LMD
Firm FEI
3006293618
Firm city
Vancouver
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register