FDA Medical Device Recalls (openFDA)
Applied Medical Resources Corp: Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments.
- Recall number
- Z-0644-2013
- Recalling firm
- Applied Medical Resources Corp
- Product
- Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments.
- Status
- Terminated
- Initiated
- 2012-11-05
- Posted
- 2013-01-07
- Terminated
- 2013-01-16
- Reason
- Applied Medical is conducting a voluntary recall on a specific lot number of its 12x150mm Kii shielded bladed access systems. This is due to a potential for the blade to remain exposed after shield deployment. Further use of any affected product should cease immediately. The affected model number is CTB71, and the affected lot is 1179107.
- Root cause
- Nonconforming Material/Component
- Action
- Applied Medical sent an Urgent Medical Device Recall to all affected customers in November 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product, complete the attached Recall Notification Confirmation Form to acknowledge the recall and indcated any product their facility is returning. The Recall confirmation form should be returned to Applied Medical by emailing to recall60494844#applied medical.com or fax to 949-713-8832. Return affecte4d product and a copy of the confirmation form to Applied Medical. Customers with questions were instructed to call 949-713-8662. For questions regarding this recall 949-713-8662.
- Quantity
- 72 units
- Distribution
- Worldwide Distribution - USA (nationwide) and France.
- Lot, serial or model codes
- Lot Number: 1179107
- 510(k) numbers
- ["K072674"]
- PMA numbers
- not on file
- Product code
- CGJ
- Firm FEI
- 2027111
- Firm city
- Rancho Santa Margarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28