Ortho-Clinical Diagnostics: VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Unique Device Identifier Number 10758750001644, Software Version 2.8 and below, IVD. Product Usage: The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.).

Recall number
Z-0643-2016
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Unique Device Identifier Number 10758750001644, Software Version 2.8 and below, IVD. Product Usage: The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.).
Status
Terminated
Initiated
2015-11-03
Posted
2016-01-13
Terminated
2018-05-30
Reason
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
Root cause
Software design
Action
Ortho-Clinical Diagnostics sent an Urgent Product Correction Letters (Ref. CL2015-211, dated November 3, 2015) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected devices (VITROS 3600 and 5600 Systems). The notification identified the affected products, problem and actions to be taken. Foreign affiliates were informed via e-mail on 11/03/2015 of the issue and instructed to notify their consignees of the issue and required actions. Ortho-Clinical Diagnostics sent an Urgent Product Correction Letters (Ref. CL2015-221, dated November 11, 2015) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected devices (VITROS 5,1 FS and 4600 users). The notification identified the affected products, problem and actions to be taken. Foreign affiliates were informed via e-mail on 11/11/2015 of the issue and instructed to notify their consignees of the issue and required actions.
Quantity
VITROS 5,1: Domestic - 932, Foreign - 1245; 5,1 Refurbished: Domestic - 63; Foreign - 153
Distribution
Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.
Lot, serial or model codes
Software Version 2.8 and below; VITROS 5,1 FS Chemistry System Serial Numbers J34000016 - J34002323; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers 34000133, 34000193, 34000197, 34000212,  34000285, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451,  34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557, 34000590, 34000596, 34000601, 34000631, 34000642, 34000761, 34000799, 34000805, 34000833, 34000851, 34000893, 34000946, 34000947, 34001006, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274,  34001297, 34001311, 34001317, 34001345, 34001378, 34001379, 34001389,  34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572,  34001612, 34001626, 34001633, 34001668, 34001755, 34001809, 34001840,  34001850, 34001875, 34001897, 34000123, 34000143, 34000191, 34000215, 34000219, 34000235, 34000251, 34000256, 34000260, 34000261, 34000319,  34000331, 34000380, 34000382, 34000406, 34000408, 34000412, 34000426, 34000431, 34000454, 34000464, 34000473, 34000501, 34000523, 34000530,  34000541, 34000542, 34000546, 34000548, 34000553, 34000558, 34000576,  34000577, 34000584, 34000585, 34000609, 34000612, 34000615, 34000616, 34000617, 34000618, 34000620, 34000625, 34000637, 34000644, 34000672,  34000744, 34000745, 34000755, 34000759, 34000776, 34000793, 34000800, 34000814, 34000831, 34000844, 34000846, 34000852, 34000856, 34000870,  34000876, 34000887, 34000892, 34000902, 34000912, 34000920, 34000938,  34000956, 34000969, 34000976, 34001007, 34001009, 34001044, 34001045,  34001052, 34001071, 34001075, 34001098, 34001099, 34001111, 34001129,  34001154, 34001161, 34001171, 34001182, 34001192, 34001195, 34001201, 34001204, 34001208, 34001212, 34001234, 34001247, 34001254, 34001271,  34001278, 34001284, 34001288, 34001289, 34001293, 34001298, 34001301,  34001302, 34001316, 34001325, 34001337, 34001340, 34001348, 34001367,  34001369, 34001399, 34001400, 34001407, 34001410, 34001416, 34001423,  34001427, 34001429, 34001457, 34001490, 34001518, 34001526, 34001551,  34001573, 34001580, 34001581, 34001582, 34001587, 34001589, 34001594,  34001610, 34001618, 34001623, 34001627, 34001673, 34001688, 34001693,  34001725, 34001738, 34001745, 34001747, 34001790, 34001796, 34001798,  34001810, 34001822, 34001831, 34001833, 34001849, 34001855, 34001859,  34001860, 34001867, 34001871, 34001872, 34001881, 34001905, 34001916,  34001923, 34001995.
510(k) numbers
["K031924","K063144"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register