Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Philips and Laerdal brands of HeartStart HS1 Defibrillator Family Model number M5066A, M5067A, and M5068A Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA. The HS1 Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The HS1 uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children.

Recall number
Z-0643-2013
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Philips and Laerdal brands of HeartStart HS1 Defibrillator Family Model number M5066A, M5067A, and M5068A Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA. The HS1 Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The HS1 uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children.
Status
Terminated
Initiated
2012-07-26
Posted
2013-01-04
Terminated
2020-09-21
Reason
Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model M5068A) automated external defibrillators (AEDs) shipped between December 2005 and July 2012. Philips had determined there is a need to emphasize the importance o
Root cause
Nonconforming Material/Component
Action
Philips Healthcare sent an Emergency Care and Resuscitation letter Customer Information Letter dated July 26, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their Philip's representative to obtain the tools and information they need to ensure that their device is functioning properly. For questions regarding this recall, please call AED Recall Support at 1-800-263-3342 (Option 6).
Quantity
Estimate 613,000 devices shipped
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Japan and Malaysia.
Lot, serial or model codes
HS1 product family all start with A in the serial number and break down into 3 catalog numbers - M5066A, M5067A and M5068A:  1) M5066A (referred to as Onsite or HS1), Philips brand, 510 (k) = K020715, sold nationally and internationally, serial numbers range from A021-00038 through A12E-03490, with 306,108 total devices.  2) M5067A, Philips brand 510(k) = K040904, sold nationally and internationally, serial numbers range from A05C-00727 through A12D-01406, with 978 total devices.  3) M5068A, Philips brand 510(k) = K020715, sold nationally and internationally, serial numbers range from A021J-00207 through A12E-01900, with 22,760 total devices   4) M5066A, Laerdal brand, 510 (k) = K020715, sold internationally, serial numbers range from A021-00036 through A12D-02148, with 60,433 total devices.  5) M5067A, Laerdal brand, sold as Home device- cleared for OTC on initial 510(k) = K040904, serial numbers range from A021-00133 through A11B-00970, with 8659 total devices.  6) M5068A, Laerdal brand, 510(k) = K020715, sold internationally, serial numbers range from A02K-00172 through A11H-00868, with 5217 total devices.  Please note the serial number range represents first and last serial number for each model. Serial numbers are not continguous.  Laerdal is a distribution partner in EU and Canada, sold internationally
510(k) numbers
["K020715"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register