Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: MEDLINE convenience kits labeled as: 1) LITHOTOMY/PELVISCOPY PACK, REF DYNJ0531985R; 2) PELVISCOPY PACK-LF, REF DYNJ0843760I; 3) PELVISCOPY PACK-LF, REF DYNJ0843760J

Recall number
Z-0642-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
MEDLINE convenience kits labeled as: 1) LITHOTOMY/PELVISCOPY PACK, REF DYNJ0531985R; 2) PELVISCOPY PACK-LF, REF DYNJ0843760I; 3) PELVISCOPY PACK-LF, REF DYNJ0843760J
Status
Open, Classified
Initiated
2024-10-09
Posted
2024-12-05
Terminated
not on file
Reason
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
Root cause
Under Investigation by firm
Action
Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email. The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal: https://recalls.medline.com. Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following: "Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-046-000, 0703-046-001, 0703-046-002. DO NOT USE THIS COMPONENT. The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*". Those further distributing the product were directed to notify their customers.
Quantity
336 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
1) REF DYNJ0531985R, UDI/DI 10195327216405 (each) 40195327216406 (case), Lot Numbers: 22HMF853, 23CMA603, 22LMG691; 2) REF DYNJ0843760I, UDI/DI 10195327365141 (each) 40195327365142 (case), Lot Numbers: 23DDA645; 3) REF DYNJ0843760J, UDI/DI 10195327491369 (each) 40195327491360 (case), Lot Numbers: 23JDA821
510(k) numbers
not on file
PMA numbers
not on file
Product code
MLT
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register