Skip to the content

FDA Medical Device Recalls (openFDA)

Orthofix, Inc: Orthofix Spinal Implants; Pillar SA PEEK Spacer System Instrument Case. Case contents: Implant Insertion Instrument, Straight Tamp, Bone Awl Drill Guide, Straight Drill, U-Joint Driver with and without retention, Flex Shaft Driver, Trial Insertion Instrument, Drill Tips, Cover Plate Inserter, Cover Plate Driver, Cover Plate Holder, Ratcheting Handle, Torque Limiting Handle, Hex Driver, Distractors/Sizers, Sizers REF: 49-0021: Orthoflex Spinal Implants, 1211 Hamburg Tnpk, Suite 300, Wayne, NJ 07470. The Pillar SA Implant Insertion Instrument is intended for intraoperative use in placing/inserting Pillar SA implants into a prepared interbody space.

Recall number
Z-0641-2010
Recalling firm
Orthofix, Inc
Product
Orthofix Spinal Implants; Pillar SA PEEK Spacer System Instrument Case. Case contents: Implant Insertion Instrument, Straight Tamp, Bone Awl Drill Guide, Straight Drill, U-Joint Driver with and without retention, Flex Shaft Driver, Trial Insertion Instrument, Drill Tips, Cover Plate Inserter, Cover Plate Driver, Cover Plate Holder, Ratcheting Handle, Torque Limiting Handle, Hex Driver, Distractors/Sizers, Sizers REF: 49-0021: Orthoflex Spinal Implants, 1211 Hamburg Tnpk, Suite 300, Wayne, NJ 07470. The Pillar SA Implant Insertion Instrument is intended for intraoperative use in placing/inserting Pillar SA implants into a prepared interbody space.
Status
Terminated
Initiated
2009-11-19
Posted
2010-01-15
Terminated
2010-02-02
Reason
The Pillar SA Implant Insertion Instrument 49-1100 may experience resistance in turning the activation knob while releasing an implant after intra-operative positioning.
Root cause
Nonconforming Material/Component
Action
A notification letter dated November 19, 2009 was sent to customers via Federal Express. The letter described the recalled product (PILLAR SA Implant Insertion Instrument, 49-1100) and problem and actions for customers to take. The affected product, PILLAR SA lmplant lnsertion lnstrument, 49-1100 is in the PILLAR SA set (set part number 49-0021) which was shipped to customers. An enclosed attachment provide instructions on product removal and return. Customers were instructed to complete the enclosed Tracking and Verification Form and fax to Orthofix Spinal Implants at 1-973-406-2912. A replacement instrument will be provided as available. In the interim, the Implant Inserter (Part Number 49-1000) included in the PILLAR SA UPE set (set number 49-0020) that was provided earlier this year may be used as an alternative to the affected product. For questions regarding the removal and return of the recalled product to Orthofix Spinal Implants, please contact your Account Services Representative. Direct other questions to the Regulatory Affairs Department at 1-973-406-2812. Direct product questions to Marketing at 1-862-221-1335 or Product Development at 1-973-406-2824.
Quantity
29 instruments
Distribution
United States (IA, NV, CO, AZ, NY, MD, IL, VA, MN, MI, NC, CT, LA, AR, GA, CA, MO, WA, OK and AL).
Lot, serial or model codes
Part number: 49-1100, Lot Number 17301-L125.
510(k) numbers
["K081849"]
PMA numbers
not on file
Product code
MQP
Firm FEI
2183449
Firm city
McKinney
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register