Zimmer Biomet, Inc.: ROSA Spine 1.0.2 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used by surgeons.

Recall number
Z-0639-2018
Recalling firm
Zimmer Biomet, Inc.
Product
ROSA Spine 1.0.2 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used by surgeons.
Status
Terminated
Initiated
2017-05-01
Posted
not on file
Terminated
2020-07-13
Reason
Robot arm being sent to the wrong position
Root cause
Software design
Action
Urgent Medical Device Recall (Correction) notices were sent to customers. The letter provided to hospital risk managers and surgeons identified the issue and their responsibilities. These responsibilities included: - Stop using the micromovement function in isocentric mode during Brain Surgery with ROSA Brain 3.0 device. -Reviewing the notification to ensure comprehension of the steps required to be taken prior to the permanent correction that will be implemented on site by a service engineer. - Completing and returning Attachment 1 (Certificate of Acknowledgement) to the e-mail address provided. The firm has prepared a software correction that will be deployed and implemented on site by a MEDTECH representative.
Quantity
1
Distribution
Distributed to CA, DC, FL, IL, KY, MA, MN, NC, NY, OH, and PA.
Lot, serial or model codes
SP16008
510(k) numbers
["K151511"]
PMA numbers
not on file
Product code
OLO
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register