FDA Medical Device Recalls (openFDA)

Haag-Streit USA Inc: Octopus 900 Perimeter, EyeSuite Perimetry version i4.000 (Flash v. 2.1.4 & 2.1.5), Model Number: Octopus 900, Distributed by: Haag-Streit USA, Inc., 3535 Kings Mills Rd., Mason, OH 45040 Product Usage: The OCTOPUS 900 Perimeter is designed for the examination, analysis and documentation of the field of sight, especially the light difference sensitivity and other functions of the human eye.

Recall number
Z-0638-2012
Recalling firm
Haag-Streit USA Inc
Product
Octopus 900 Perimeter, EyeSuite Perimetry version i4.000 (Flash v. 2.1.4 & 2.1.5), Model Number: Octopus 900, Distributed by: Haag-Streit USA, Inc., 3535 Kings Mills Rd., Mason, OH 45040 Product Usage: The OCTOPUS 900 Perimeter is designed for the examination, analysis and documentation of the field of sight, especially the light difference sensitivity and other functions of the human eye.
Status
Terminated
Initiated
2011-11-14
Posted
2012-01-11
Terminated
2014-06-26
Reason
Currently there is a possibility, at the start of the perimetry examination, for the background illumination of the cupola not to turn on. If no illumination of cupola occurs, data obtained from the examination could provide the doctor with results that would appear to be better than actual.
Root cause
Software design
Action
Haag-Streit USA Inc sent an Urgent Field Safety Notice letter dated November 11, 2011 to their customers. The letter identified the product, the problem, and the actions to be taken. Customers were instructed to check their unit and if needed upgrade their unit with the new software provided. Customers were advised to complete and return the Software Update Verification Form included. Customers who had distributed the affected product outside of their facility were instructed to immediately notify their customers of the recall.
Quantity
158 Units
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA and WA and the country of Mexico.
Lot, serial or model codes
Model Number: Octopus 900, S/N's: 159, 231, 279, 313, 314, 315, 336, 338, 478, 479, 484, 485, 559, 560,561, 562, 563, 642, 644, 645, 646, 647, 648, 649, 650, 651, 772, 773, 774, 775, 776, 777, 797, 798, 804, 805, 806, 807, 808, 813, 814, 839, 840, 841, 842, 843, 844, 868, 869, 870, 871, 872, 873, 874, 903, 904, 905, 907, 908, 909, 910, 911, 956, 957, 958, 959, 975, 983, 984, 985, 986, 987, 1002, 1003, 1004, 1005, 1006, 1007, 1008, 1034, 1035, 1036, 1037, 1038, 1059, 1060, 1061, 1062, 1063, 1064, 1065, 1066, 1117, 1118, 1119, 1120, 1140, 1141, 1143, 1152, 1162, 1163, 1164, 1166, 1199, 1200, 1201, 1202, 1239, 1240, 1241, 1262, 1263, 1286, 1287, 1288, 1307, 1308, 1401, 1402, 1403, 1404, 1405, 1424, 1425, 1426, 1458, 1459, 1460, 1480, 1481, 1482, 1483, 1484, 1543, 1545, 1546, 1550, 1552, 1553, 1554, 1555, 1556, 1560, 1561, 1562, 1563, 1564, 1579, 1580, 1581, 1582, 1583, 1586, 1587, 1588, 1589 and 1591.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HJB
Firm FEI
1000136533
Firm city
Mason
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register