Skip to the content

FDA Medical Device Recalls (openFDA)

Kowa Optimed Inc: Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for acquiring, displaying and storing image data from KOWA s mydriatic and non-mydriatic ophthalmic cameras.

Recall number
Z-0636-2021
Recalling firm
Kowa Optimed Inc
Product
Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for acquiring, displaying and storing image data from KOWA s mydriatic and non-mydriatic ophthalmic cameras.
Status
Terminated
Initiated
2020-11-03
Posted
not on file
Terminated
2022-12-06
Reason
It has been determined that the software distributed with the retinal camera may delete patient information. Specifically, if multiple IDs are delated at the same time, their is a potential for all patient information to be deleted. This could result in delay of treatment.
Root cause
Under Investigation by firm
Action
On 11/03/2020 Kowa Company mailed a "Urgent Medical Device Recall letter to all affected consignees. In addition to informing consignees about the recalled software, the firm asked consignees to take the following actions: 1. The service team at Kowa will contact each customer and remove version 10.0.4.0. on all workstations and provide version 10.0.4.100. A CD of version 10.0.4.100 will be mailed to each customer, but the removal and software update will be performed remotely. Other Information: 2. Complete the Medical Device Acknowledgement and receipt form and fax completed response form to : 310-327-4177 ATTN: Benjamin Rhodes or mail to: Kowa American Corporation, 20001 South Vermont Avenue, Torrance, CA 90502. 3. " For more information, please contact Benjamin Rhodes at brhodes@kowa.com " For specific information regarding the remote software update, please contact service@kowa.com
Quantity
not on file
Distribution
US Nationwide distribution including in the states of NY, WA, TX, CT, MA, ID, IL, MI.
Lot, serial or model codes
Serial Numbers: 32168900007, 32168900008, 32168900009, 32168900010, 32168900011, 32168900019, 32168900020, 32168900021, 32168900022, 32168900025, 32168900035, 32168900036, 32168900037, 32168900044, 32168900045, 32169000052, 32169000144, 32169000145, 32169000152, 32169000153, 32169000166, 32169000167, 32169000168
510(k) numbers
["K190056"]
PMA numbers
not on file
Product code
NFJ
Firm FEI
2080939
Firm city
Torrance
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register