Skip to the content

FDA Medical Device Recalls (openFDA)

Helena Laboratories, Corp.: K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING TIME ANALYZER, and ACTALYKE MINI ACTIVATED CLOTTING TIME TEST SYSTEM.

Recall number
Z-0635-2021
Recalling firm
Helena Laboratories, Corp.
Product
K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING TIME ANALYZER, and ACTALYKE MINI ACTIVATED CLOTTING TIME TEST SYSTEM.
Status
Open, Classified
Initiated
2020-11-09
Posted
not on file
Terminated
not on file
Reason
Due to complaints received associated with cracked/split test tube caps.
Root cause
Nonconforming Material/Component
Action
On 11/09/2020, international distributors were initially contacted and provided with a notification for distributors letter. On 11/16/2020 domestic customers were contacted via telephone and then a formal notification was delivered. The Recalling Firm was informing its customers to the potential of defect associated with the (flip-top) caps for some test tubes where the cap my split resulting in leaking of biohazardous fluids as users are instructed to shake the tube for proper mixing. Customers are instructed to immediately examine their inventory and quarantine affected product. If customers have further distributed the affected product, they are asked to identify the users within their facility to notify them at once of this product recall. Customers are also asked to complete and return the response form as soon as possible. For any questions, customer can email or call the Recalling Firm as jhanka@Helena.com or 1-800-231-5663 extension 1177.
Quantity
Lot 1-19-K-ACT = 314 tubes and Lot 2-19-K-ACT = 297 tubes
Distribution
Worldwide distribution - U.S. Nationwide including in the states of NJ, NY, PA and TX. The countries of Austria, Canada, England, France, Italy, Portugal, Russia, Spain.
Lot, serial or model codes
Catalog Number: K-ACT Lot Number: 1-19-K-ACT and 2-19-K-ACT
510(k) numbers
["K013078","K061014"]
PMA numbers
not on file
Product code
JPA
Firm FEI
1618982
Firm city
Beaumont
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register