Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics; UHR Universal Bipolar Component; 26 x 50 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430

Recall number
Z-0635-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics; UHR Universal Bipolar Component; 26 x 50 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430
Status
Terminated
Initiated
2007-10-26
Posted
2008-03-20
Terminated
2010-01-27
Reason
Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head.
Root cause
Process control
Action
Stryker sent Urgent Product Hold letters to US branches/agencies and hospitals on October 26, 2007 advising them to quarantine the product from these lots, and Urgent Product Recall letters on November 2, 2007 advising them to return the devices to Stryker. International distributors were sent an advisory on November 6, 2007.
Quantity
360 total
Distribution
Worldwide Distribution - USA, Canada, China, India, Australia, Korea, Singapore, Taiwan, Sweden, Germany, Netherlands, France, Spain, Italy, Poland, South Africa, and England.
Lot, serial or model codes
Ref No: Description: Lot Code:  UH1-50-26 UHR BIPOLAR 26X50mm 6X5MAD UH1-50-26 UHR BIPOLAR 26X50mm J3VMKA UH1-50-26 UHR BIPOLAR 26X50mm J84MDA UH1-50-26 KJ1MEA UH1-50-26 KJ4MEA UH1-50-26 KJ8MEA UH1-50-26 KJ9MEA UH1-50-26 KJAMEA
510(k) numbers
["K800207"]
PMA numbers
not on file
Product code
KWL
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register