Skip to the content

FDA Medical Device Recalls (openFDA)

Horiba Instruments, Inc dba Horiba Medical: PENTRA C200, Part Number: A11A01629. The Pentra C200 system is a fully automated chemistry analyzer using colorimetry, turbidmetry and potentiometry technologies. It is mostly dedicated to in vitro diagnostic analyses based on homogeneous samples such as serum, plasma and uring.

Recall number
Z-0633-2013
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Product
PENTRA C200, Part Number: A11A01629. The Pentra C200 system is a fully automated chemistry analyzer using colorimetry, turbidmetry and potentiometry technologies. It is mostly dedicated to in vitro diagnostic analyses based on homogeneous samples such as serum, plasma and uring.
Status
Terminated
Initiated
2012-10-10
Posted
2013-01-03
Terminated
2013-09-12
Reason
Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
Root cause
Software design
Action
Horiba Medical sent a Field Safety Notification and Acknowledgement Letter dated November 20, 2012, to all affected customers. The letter notified the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 calendar days via mail, email or fax. Customers with questions were instructed to contact their local representative. For questions regarding this recall call 714-273-9254.
Quantity
384 units
Distribution
Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX.
Lot, serial or model codes
Lot number: 00456-4108
510(k) numbers
["K103788"]
PMA numbers
not on file
Product code
JJE
Firm FEI
3002698356
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register