Skip to the content

FDA Medical Device Recalls (openFDA)

Diasorin Inc.: ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285

Recall number
Z-0633-05
Recalling firm
Diasorin Inc.
Product
ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285
Status
Terminated
Initiated
2004-10-20
Posted
2005-03-18
Terminated
2006-07-22
Reason
2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.
Root cause
Other
Action
Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.
Quantity
not on file
Distribution
MA, OH & IL
Lot, serial or model codes
lot # E052660D, E052790D
510(k) numbers
not on file
PMA numbers
not on file
Product code
LAZ
Firm FEI
2182595
Firm city
Stillwater
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register