FDA Medical Device Recalls (openFDA)
Diasorin Inc.: ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285
- Recall number
- Z-0633-05
- Recalling firm
- Diasorin Inc.
- Product
- ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285
- Status
- Terminated
- Initiated
- 2004-10-20
- Posted
- 2005-03-18
- Terminated
- 2006-07-22
- Reason
- 2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.
- Root cause
- Other
- Action
- Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.
- Quantity
- not on file
- Distribution
- MA, OH & IL
- Lot, serial or model codes
- lot # E052660D, E052790D
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LAZ
- Firm FEI
- 2182595
- Firm city
- Stillwater
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28