Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems Inc: VARiS 1.4G and Vision/SomaVision Radiation Therapy Systems; VARiS versions 6.1.33 and 6.2.19 in conjunction with Eclipse 6.x/Soma Vision 6.x

Recall number
Z-0633-03
Recalling firm
Varian Medical Systems Inc
Product
VARiS 1.4G and Vision/SomaVision Radiation Therapy Systems; VARiS versions 6.1.33 and 6.2.19 in conjunction with Eclipse 6.x/Soma Vision 6.x
Status
Terminated
Initiated
2003-01-31
Posted
2003-03-12
Terminated
2003-09-17
Reason
The product has a software problem. This could result in a treatment without a motorized wedge where one had been planned.
Root cause
Software design
Action
On 1/16/03 and 2/4/03 (2nd notification) the firm initiated the recall and their notification was via letters to inform its consignees about the new VARiS version 6.2.27 release.
Quantity
12 units
Distribution
Product was released for distribution to 7 hospitals in the US. Distribution was limited to the following States: CA, MA, NC, SC, PA, and LA. 1 (one) Canadian distribution includes Fraser Valley Cancer Center, Dept. of Radiation Oncology, 13750 96th Ave., Surrey, BC, V3V 1Z2, Canada. Foreign distribution includes Austria, Germany and Finland. The recall was appropriately extended to the user level; i.e., the hospitals/physicians who received the recalled product.
Lot, serial or model codes
Affected Serial Numbers: 44, 260, 315, 495, 746, 751, 873, 907, 984, 1098, 1131, and 9045
510(k) numbers
["K001643"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register