Skip to the content

FDA Medical Device Recalls (openFDA)

Perkinelmer: Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Recall number
Z-0632-2016
Recalling firm
Perkinelmer
Product
Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
Status
Terminated
Initiated
2015-12-08
Posted
2016-01-12
Terminated
2017-08-04
Reason
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
Root cause
Nonconforming Material/Component
Action
Perkin Elmer sent an Urgent Medical Device Correction letter dated December 7, 2015 to affected customers. The letter identified the affected products, problems and actions to be taken. Customers were instructed to destroy all affected products in their possession. Customers were asked to complete the enclosed Correction Response Form and return it by scan and e-mail to QMresponse.SG@perkinelmer.com . Corrected barcode labels ID #023 will be provided upon request and return of the completed Correction Response Form to PerkinElmer. For further information contact your local PerkinElmer representatives..
Quantity
47 units
Distribution
Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.
Lot, serial or model codes
Serial Numbers: DG06106539 DG12118720 DG07106646 DG12107250 DG12107281 DG03117619 DG11118605 DG01117400 DG12107202 DG01117401 DG06106449 DG03117666 DG05117791 DG06106454 DG03106180 DG11118637 DG07118121 DG07106598 DG11118633 DG06106520 DG03106137 DG07118111 DG02128888 DG08118247 DG06118014 DG01128790 DG12107207 DG07106637 DG01128876 DG04117764 DG01106035 DG12118780 DG05106342 DG01117399 DG06117909 DG06118011 DG09106846 DG11107146 DG12118694 DG08118217 DG02106084 DG04117690 DG05117838 DG09106799 DG04117735 DG03117548 DG01117362
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJJ
Firm FEI
3006159931
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.