Gyrus ACMI Corporation: Gyrus ACMI PKS Seal Open Forceps, model 3103PK, 9-3/4" curved. Sterile EO. The device is intended to be used with the PlasmaKinetic electrosurgical generator, software version 1.61 or greater required or the PlasmaKinetic with SuperPulse electrosurgical generator, software version 2.0 or greater or G400 electrosurgical generator. Gyrus ACMI, Inc., 136 Turnpike Road, Southborough, MA 01772-2104 USA.

Recall number
Z-0632-2010
Recalling firm
Gyrus ACMI Corporation
Product
Gyrus ACMI PKS Seal Open Forceps, model 3103PK, 9-3/4" curved. Sterile EO. The device is intended to be used with the PlasmaKinetic electrosurgical generator, software version 1.61 or greater required or the PlasmaKinetic with SuperPulse electrosurgical generator, software version 2.0 or greater or G400 electrosurgical generator. Gyrus ACMI, Inc., 136 Turnpike Road, Southborough, MA 01772-2104 USA.
Status
Terminated
Initiated
2009-08-18
Posted
2010-01-14
Terminated
2012-03-06
Reason
Gyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field. The detached shim may be undetected and be left behind in the patient. This circumstance could subsequently cause complications. Further use of this product should cease, affected product should be removed from stockrooms and returned to Gyrus ACMI.
Root cause
Environmental control
Action
An Urgent Medical Device Recall letter, dated 08/15/09 was sent to customers notifying them of the issue. The letter identified affected product, asked that they cease further use, remove affected devices from stockrooms, contact Gyrus ACMI customer service for a return goods authorization and return product to Gyrus ACMI at no charge. A reply form is also asked to be returned to Gyrus ACMI via fax. Questions and requests for a Return Goods Authorization are directed to Gyrus ACMI customer service at 1-888-524-7266.
Quantity
93,665 total (63,418 US, 30,247 OUS)
Distribution
US and International distribution. OUS to Include: Australia, Belgium, Bolivia, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece , Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Panama, Puerto Rico, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Lot, serial or model codes
all lots
510(k) numbers
["K024286"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3003790304
Firm city
Southborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register