Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien: Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD

Recall number
Z-0631-2024
Recalling firm
Covidien
Product
Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD
Status
Open, Classified
Initiated
2023-11-16
Posted
2024-01-03
Terminated
not on file
Reason
Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
Root cause
Software Manufacturing/Software Deployment
Action
On November 16 and 20, 2023, Covidien LLC (Medtronic Company) issued a "Urgent: Medical Device Correction" 1. Immediately notify all personnel in all care environments in which the Valleylab" FT10 Energy Platform. 2. Update ValleylabTM FT10 Energy Platform to software version 4.0.4. For additional information see customer communication letter 3. Until the software is updated, the Valleylab" FT10 Energy Platform and LigaSure" devices can continue to be used as instructed in the User Guide and per your facility protocols. 4. Complete the attached Customer Confirmation Form and return it as directed to confirm your receipt and understanding of this information.
Quantity
47,901 units
Distribution
Worldwide distribution - US: Nationwide and surrounding territories.
Lot, serial or model codes
Ref-UDI: VLFT10GEN- 10884521516328, DLVLFT10GEN - 10884521787056 , VLFT10GENZD - 10884521640474/Software: v4.0.1, v4.0.2 and v4.0.3
510(k) numbers
["K191601"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3026612481
Firm city
Lafayette
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register