FDA Medical Device Recalls (openFDA)
Arkray Factory USA, Inc.: arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
- Recall number
- Z-0631-2016
- Recalling firm
- Arkray Factory USA, Inc.
- Product
- arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
- Status
- Terminated
- Initiated
- 2015-12-18
- Posted
- 2016-01-15
- Terminated
- 2017-01-04
- Reason
- One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
- Root cause
- Component design/selection
- Action
- The firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243.
- Quantity
- 11 boxes (25 foil packaged test strips per box)
- Distribution
- Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
- Lot, serial or model codes
- Lot Number PN5C26
- 510(k) numbers
- ["K053401"]
- PMA numbers
- not on file
- Product code
- CGA
- Firm FEI
- 1832816
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28