FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx only, STERILE
- Recall number
- Z-0630-2014
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx only, STERILE
- Status
- Terminated
- Initiated
- 2013-12-06
- Posted
- 2014-01-06
- Terminated
- 2014-02-27
- Reason
- The Color Cuff Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector. A customer reported having received a luer connector instead of a quick connect fitting. An incorrect connector may lead to an inoperable system due to incompatibility with the Smart Pump.
- Root cause
- Component change control
- Action
- Distributor was notified by E-mail 11/26/2013, with follow-up UPS letters sent to Hospital Risk/ Materials Managers identifying the product, reason for the recall, and risk. The notice asked consignees to find and quarantine all implicated product, with hospitals requested to complete the Business Reply Form and FAX to 866-521-2762. Stryker will email a pre-paid shipper and credit all accounts. Question and concerns should be addressed to Kelly Jo Whipple 269-389-2921 or kellyjo.whipple@stryker.com
- Quantity
- 5 boxes (50 units)
- Distribution
- US Distribution: including states of: MI, OH, and TX.
- Lot, serial or model codes
- Lot 50731155 Color code Purple
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KCY
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28