Skip to the content

FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc. d.b.a. Mindray North America: DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring Station (CMS) network transfers information between DOM Central Station and other network devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and or WLAN connections. The DPM Central Station is used for remote monitor management, storing, printing, reviewing, or processing of information from net worked devices, and it is operated by medical personnel in hospitals or medical institutions.

Recall number
Z-0630-2013
Recalling firm
Mindray DS USA, Inc. d.b.a. Mindray North America
Product
DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring Station (CMS) network transfers information between DOM Central Station and other network devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and or WLAN connections. The DPM Central Station is used for remote monitor management, storing, printing, reviewing, or processing of information from net worked devices, and it is operated by medical personnel in hospitals or medical institutions.
Status
Terminated
Initiated
2011-05-12
Posted
2013-01-03
Terminated
2013-02-19
Reason
There is a software anomaly with the DPM Central Monitoring System where the trend data of one patient maybe replaced with the trend data of a second patient.
Root cause
Software design
Action
Mindray sent a Product Corrective Action letter dated May 12, 2012, to all affected customers via certified mail return receipt requested to two medical facilities. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that the issue would be corrected through a softweare upgrade of the affected product. A Mindray Service Representative would contact their facility to arrange for the software upgrade at no cost to them. For questions regarding this recall call 1-800-288-2121.
Quantity
two units
Distribution
Nationwide Distribution including OH and GA
Lot, serial or model codes
510 K K080192 P/N 300BF-PA200002
510(k) numbers
["K080192"]
PMA numbers
not on file
Product code
MSX
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register