Skip to the content

FDA Medical Device Recalls (openFDA)

Materialise USA LLC: REF 42-411571 QTY 1 CATALOG CAUTION. SEE INSTRUCTIONS FOR USE SIGNATURE M PKA GDE/MDL SET SML-MP DD.....R-FEB55RKA-T JA-A01 05551 PKA POST CUT SML-PKA MICROPLASTY CUSTOM-MADE DEVICE non-sterile The Signature Personalized Patient Care System includes the patient-specific Signature Guides which are patient specific instruments (guides are considered custom-made per EU definition) intended for use during the surgical procedure for total and partial knee arthroplasty.

Recall number
Z-0628-2014
Recalling firm
Materialise USA LLC
Product
REF 42-411571 QTY 1 CATALOG CAUTION. SEE INSTRUCTIONS FOR USE SIGNATURE M PKA GDE/MDL SET SML-MP DD.....R-FEB55RKA-T JA-A01 05551 PKA POST CUT SML-PKA MICROPLASTY CUSTOM-MADE DEVICE non-sterile The Signature Personalized Patient Care System includes the patient-specific Signature Guides which are patient specific instruments (guides are considered custom-made per EU definition) intended for use during the surgical procedure for total and partial knee arthroplasty.
Status
Terminated
Initiated
2013-11-26
Posted
2014-01-03
Terminated
2014-01-14
Reason
Shipped femur guide contained a non-conformance which could possibly impact fit of the drill insert.
Root cause
Nonconforming Material/Component
Action
Materialise USA LLC notified the Field Representative on November 26, 2013 of the recall by phone followed by E-mail notification. The notification identified the device and reason for the recall. Notification directions included instructions to find and quarantine the device and return it Materialise. A replacement device was shipped November 27, 2013 receiving the defective device was called, followed by E-mail notification the Recall, which identified the device, problem, and instructed the distributor to find and quarantine product. For questions regarding this recall call 734-662-5057.
Quantity
1
Distribution
US Distribution: MO only.
Lot, serial or model codes
Lot 094929
510(k) numbers
["K110415"]
PMA numbers
not on file
Product code
HRY
Firm FEI
3005718816
Firm city
Plymouth
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28