Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: RINGLOC ACETABULAR SYSTEM Self-tapping bone screws, Item numbers 103531, 103533, 103534, 103535 - Product Usage: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented; Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented; Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component); Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer.

Recall number
Z-0627-2021
Recalling firm
Biomet, Inc.
Product
RINGLOC ACETABULAR SYSTEM Self-tapping bone screws, Item numbers 103531, 103533, 103534, 103535 - Product Usage: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented; Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented; Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component); Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer.
Status
Terminated
Initiated
2020-11-25
Posted
2020-12-17
Terminated
2024-05-08
Reason
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
Root cause
Process control
Action
On November 25, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed of the potential loss of sterility in affected devices. Customers were asked to do the following: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. All customers should also complete the certificate of acknowledgement included with the letter and return to the firm. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity
7428
Distribution
Worldwide distribution. US nationwide, ARGENTINA, AUSTRALIA, Brazil, CANADA, CHILE, CHINA, COSTA RICA, EMEA, HONG KONG, INDIA, JAPAN, KOREA, MALAYSIA, NEW ZEALAND, SINGAPORE, and TAIWAN
Lot, serial or model codes
Lots 259830 259860 259880 259890 259910 259920 260020 260140 260160
510(k) numbers
["K063570","K082446","K912245","K971890","K990032"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register