Zimmer Biomet, Inc.: AGC 2000 LP Patella, labeled as AGC Porous Patellar SM, Model Number 150802. Product Usage: For use in Total Knee Arthroplasty.

Recall number
Z-0627-2018
Recalling firm
Zimmer Biomet, Inc.
Product
AGC 2000 LP Patella, labeled as AGC Porous Patellar SM, Model Number 150802. Product Usage: For use in Total Knee Arthroplasty.
Status
Terminated
Initiated
2017-08-25
Posted
not on file
Terminated
2018-08-09
Reason
The affected AGC 2000 LP Patella products were over-molded leaving the outside diameter nonconforming to the print dimension.
Root cause
Process change control
Action
Consignees were notified via mail on approximately 08/25/2017. Instructions to distributors, sales representatives, and distributor operations managers include to ensure affected team members are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide any additional hospitals and/or surgeons that received direct shipments. Instructions to risk managers and surgeons include to ensure affected personnel are informed of the recall, quarantine all affected product, return any affected product in inventory, and complete and return the Certificate of Acknowledgment. Questions or concerns can be sent to the Zimmer Biomet customer call center at 1-574-371-3071 between 8:00 am and 5:00 pm EST, Monday through Friday. Or they can be sent via email to corporatequality.postmarket@zimmerbiomet.com.
Quantity
28 devices
Distribution
US Nationwide Distribution to IN, MO, NY, TX.
Lot, serial or model codes
Model No. 150802: Lot No. 305330 (UDI (01)00880304240230(17)220506(10)305330), 305340 ((01)00880304240230(17)220506(10)305330), 305350 ((01)00880304240230(17)220517(10)305350), 356750 ((01)00880304240230(17)220523(10)356750).
510(k) numbers
["K833921"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register