FDA Medical Device Recalls (openFDA)
Randox Laboratories Ltd.: SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349
- Recall number
- Z-0624-2024
- Recalling firm
- Randox Laboratories Ltd.
- Product
- SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349
- Status
- Open, Classified
- Initiated
- 2023-11-06
- Posted
- 2024-01-02
- Terminated
- not on file
- Reason
- A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
- Root cause
- Process control
- Action
- Randox issued Urgent Medical Device Correction Letter on 11/6/23 to the US Distributor Randox Laboratories-US, Ltd. , Kearneysville, West Virginia, for issuance to the user level. Letter states reason for recall, health risk and action to take: As the control material is still testing positive for HSV Type 1/2 lgM, there is no risk to health. If the control did test negative for this analyte, the affected run should be discarded, and the samples re-analysed. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. " Please discard all copies of the IFU and download the latest version from www.randox.com.alth risk and action to take: (
- Quantity
- 1 unit
- Distribution
- NY, WV
- Lot, serial or model codes
- GTIN: 05055273216424 Bath/Lot Number: 157SR
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OHQ
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28