Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349

Recall number
Z-0624-2024
Recalling firm
Randox Laboratories Ltd.
Product
SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349
Status
Open, Classified
Initiated
2023-11-06
Posted
2024-01-02
Terminated
not on file
Reason
A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
Root cause
Process control
Action
Randox issued Urgent Medical Device Correction Letter on 11/6/23 to the US Distributor Randox Laboratories-US, Ltd. , Kearneysville, West Virginia, for issuance to the user level. Letter states reason for recall, health risk and action to take: As the control material is still testing positive for HSV Type 1/2 lgM, there is no risk to health. If the control did test negative for this analyte, the affected run should be discarded, and the samples re-analysed. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. " Please discard all copies of the IFU and download the latest version from www.randox.com.alth risk and action to take: (
Quantity
1 unit
Distribution
NY, WV
Lot, serial or model codes
GTIN: 05055273216424 Bath/Lot Number: 157SR
510(k) numbers
not on file
PMA numbers
not on file
Product code
OHQ
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register